VSWR Measurement and Set-up for measurement
The following steps will configure the Network Analyzer with an
"Adapter Removal" Calibration for measurement of the window..
Cal Kit Set-up:
- Two Cal Kits are used. One Cal Kit is the WG16x9 which has the
"adapter" information. The other Cal Kit is the WR2100_86
when using the 8.6" delay line of WR2100.
- These Cal Kits need to be loaded into the Cal 1 and the Cal 2 storage
locations under the "Calibration Menu" on the Network Analyzer.
- Remember to set the Impedance to 1 ohm for the waveguide calibration.
- Make sure there are three Cal Set storage locations before starting the
calibration procedure.
Stimulus:
- Center Frequency 476 MHz
- Span 100 MHz
Calibration:
- Perform a TRL calibration for Port 1 and Port 2 with the
"Adapter" attached to Port 1 and Port 2 using the Cal Kit for the
waveguide size of Port 1 and Port 2.
- Store the Cal Set of Port 1 in storage location 1 and the Cal Set for Port
2 in storage location 2
- The following steps are now a sequence of instructions using the buttons
and questions on the screen of the network analyzer:
-
MORE
-
MODIFY CAL SET
-
ADAPTER REMOVAL
-
CAL SET FOR PORT 1 : CAL SET 1
-
CAL SET FOR PORT 2 : CAL SET 2
-
ADAPTER WG16x9
-
MODIFY AND SAVE : CAL SET 3
For the "thru" portion of the calibration
the set-up looks like this:

When doing the WR2100
Calibration, the set-up looks like this: (The adapter is part of the WG16x9
port.)

When doing the WG16x9
Calibration the set-up looks like this: (The adapter is now part of the WR2100
port.)